1. General Introduction
ABB EL3000 combine different analyzers in one single housing provides optimum economic and operational efficiency for your application. Different types of detectors offers multiple process and emission monitoring, especially for measuring flammable gases, five sample components cam be integrated each device. Similar as AO2000,it compose of sampling system, Analyzer modules, power supply, I/O modules and software packages. Two type of housing available, EL3020 is the IP20 19-inch rack mounting type while EL3040 is IP65 wall-mount type. Five of analyzers modules can be combined from bellow:
l Uras26 Infrared photometer
l Limas23 Ultraviolet photometer
l Caldos27 Thermal conductivity analyzer
l Magnos28 Paramagnetic oxygen analyzer
l Magnos206 Oxygen analyzer
l ZO23 Trace Oxygen analyzer
l Fidas24 Flame Ionization Detector
2. Typical Sample components
Analyzer Module & principle | Sample | Principle | Smallest Range | Largest Range |
Infrared Photometer Uras26 | CO | Non-dispersive infrared absorption | 0…100 ppm | 0…500 ppm |
CO2 | Non-dispersive infrared absorption | 0…100 ppm | 0…500 ppm | |
NO | Non-dispersive infrared absorption | 0…150 ppm | 0…750 ppm | |
SO2 | Non-dispersive infrared absorption | 0…100 ppm | 0…500 ppm | |
N2O | Non-dispersive infrared absorption | 0…100 ppm | 0…500 ppm | |
CH4 | Non-dispersive infrared absorption | 0…100 ppm | 0…500 ppm | |
Ultraviolet photometer Limas23 | NO | UV Resonance Absorption Spectroscopy | 0…50 ppm | 0…5000 ppm |
NO2 | Non-Dispersive UV Absorption Spectroscopy | 0…50 ppm | 0…500 ppm | |
SO2 | Non-Dispersive UV Absorption Spectroscopy | 0…100 ppm | 0…5 vol.% | |
Oxygen analyzer Magnos206 | O2 | Paramagnetic Magneto mechanical | 0…2 vol.% | 0…100 vol.% |
Oxygen analyzer Magnos28 | O2 | Paramagnetic Magneto mechanical | 0…0.5 vol.% | 0…100 vol.% |
Oxygen analyzer Magnos27 | O2 | Paramagnetic Heavy-duty thermo magnetic | 0…10 vol.% | 0…100 vol.% |
Trace oxygen analyzer ZO23 | O2 | zirconium dioxide cell | 0…1 ppm | 0…250000 ppm |
Thermal conductivity analyzer Caldos27 | Multi | Thermal conductivity | 0…1 vol.% | 0…100 vol.% |
Flame-ionization detector Fidas24 | HC | Flame-ionization detector | 0…10 ppm | 0…10000 ppm |
3. Schematic Dimension of EL3000 housing (3020 &3040)
Address: P.O. Box 17033, Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
Tel: +971558915937
Fax: +971558915956
E-mail:
info@rostockgroup.com
technical@rostockgroup.com
commercial@rostockgroup.com
Get the latest special offers and discount information sent to your email address.
Copyright © Rostock Group Co.,Limited All Rights Reserved. Sitemap